AK 98, 115V BIO/ AK 98, 115V SELF-CARE and AK98 V2 115V BIO VERSION / AK 98 V2 115V BIO SELF CARE/ AK 98 V3 (2018-08-29)

Issued March 26, 2024
Issue
Baxter is communicating important safety information regarding the potential for AK 98 hemodialysis devices to generate excessive Ultrafiltration (UF) in certain situations where treatment-related alarms occur, or where there is an ultrafilter leak. Excessive UF may present hazards for sensitive patients, such as low-weight patients, for whom target UF values of zero or very low volumes are desired. Recall start date: August 29, 2018
Hazard description
Performance
Product
Product name
AK 98, 115V BIO/ AK 98, 115V SELF-CARE and AK98 V2 115V BIO VERSION / AK 98 V2 115V BIO SELF CARE/ AK 98 V3 (2018-08-29)
Reference
ID
RA-58388