Back to RecallTracker
Health Canada Medical devices

IntelePACS

Issued November 12, 2021
Issue
The issue may be triggered by the presence of pre-formatted html text in HL7 reports using the tag. These reports will be truncated when they are rendered by the enhanced viewer pdf report viewer, which could lead to incomplete report content displayed to the enhanced viewer user. This issue will only happen while using Microsoft Edge/Chromium 92, Mozilla Firefox 90, Google Chrome 92 and later versions of these browsers. Recall start date: 2021-10-29
Hazard description
Software defect
What to do
Contact the manufacturer if you need additional information.
Product
Product name
IntelePACS
Reference
ID
RA-63598