The affected vitros CKMB slides coating does not adequately inhibit CK-mm up to a total CK more than 1000 u/L. Therefore, samples with a total CK value less than 1000 u/L may exhibit falsely elevated vitros CKMB slides results. Recall start date: Nov 30, 2021