Back to RecallTracker
Health Canada Medical devices

Bd Veritor Plus Analyzer

Issued January 17, 2022
Issue
Bd has received two (2) complaints reporting that the Bd Veritor Plus analyzer may have the potential to overheat and/or cause fire. Bd's investigation has found that the issue can occur when an unauthorized alternative ac power adapter, and not the one provided by Bd with the Bd Veritor Plus analyzer, is used to power the instrument. Recall start date: Jan 5, 2022
Hazard description
Increased risk or new adverse events
Product
Product name
Bd Veritor Plus Analyzer
Reference
ID
RA-63828