Health Canada Medical devices

Pack,Hot/Cold,Curad,Retail W/Cov

Issued January 24, 2022
Issue
The Curad Hot/Cold gel pack with wrap currently has the labeling states "does not contain natural rubber latex". However, the fabric around the gel pack does contain some natural rubber latex. The fabric wrap contains 20% latex thread. There is a risk of using the device on an individual with latex sensitivity or allergy. Recall start date: Jan 12, 2022
Hazard description
Labelling and packaging
Product
Product name
Pack,Hot/Cold,Curad,Retail W/Cov
Reference
ID