Back to RecallTracker
Health Canada Medical devices

Acuson Juniper

Issued January 31, 2022
Issue
Siemens Healthcare is notifying customer of a potential failure in the clip store function on the Acuson Juniper ultrasound system. In Juniper 1.0 the clip store function does not work when the ultrasound system has a disk full error. The purpose of this update is to provide juniper VA10G software to all juniper systems currently installed with software versions VA10D, VA10E and VA10F as a solution. Recall start date: Jan 25, 2022
Hazard description
Performance
Product
Product name
Acuson Juniper
Reference
ID
RA-63877