Back to RecallTracker
Health Canada Medical devices

Welch Allyn Diagnostic Cardiology Products

Issued February 14, 2022
Issue
Hillrom has become aware of a vulnerability which allows the entry of any username that is provisioned in the application without supplying a password. This will then grant access to the application with the privileges of the username provided. the vulnerability occurs when customer activates single sign-on (SSO) feature in either stand alone, or client/server configurations. Recall start date: Jan 26, 2022
Hazard description
Software defect
Product
Product name
Welch Allyn Diagnostic Cardiology Products
Reference
ID
RA-63925