The recall is being initiated due to a software defect that is triggered by disconnecting / connecting an USB device, causing a restart of the infinity central station and a temporary loss of central monitoring. Recall start date: Mar 8, 2022
Hazard description
Software defect
Product
Product name
Infinity Centralstation Wide
Reference
ID
RA-63997