Back to RecallTracker
Health Canada Medical devices

Dragonfly Opstar Imaging Catheter

Issued April 25, 2022
Issue
The proximal marker on devices from these lots may separate from the device. A dislodged marker may require additional intervention, including unplanned additional coronary intervention, or surgery. To date Abbott has received 5 complaints related to this issue, one which resulted in an embolism which required an additional intervention to resolve with no immediate adverse consequences. Potential risks include embolism, cardiovascular injury and myocardial ischemia. Recall start date: Apr 11, 2022
Hazard description
Quality
Product
Product name
Dragonfly Opstar Imaging Catheter
Reference
ID