Back to RecallTracker
Health Canada Medical devices

Reliance Synergy Washer/Disinfector

Issued April 25, 2022
Issue
Steris has identified that in the remote occurrence in which the electrical contactor component present in the drying chamber of the Reliance Synergy Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire. This could result in smoke inhalation or a burn which may require medical intervention. Recall start date: Apr 14, 2022
Hazard description
Electrical defect
Product
Product name
Reliance Synergy Washer/Disinfector
Reference
ID
RA-64112