Back to RecallTracker
Health Canada Medical devices

Pagewriter Electrocardiograph

Issued May 30, 2022
Issue
The pagewriter tc30/50/70 cardiograph device could potentially fail to charge while plugged into the ac power outlet. This is due to a faulty mainboard. based on a supplier analysis, the faulty mainboard was caused by delamination of an ic chip after mold assembly. Recall start date: May 19, 2022
Hazard description
Performance
Product
Product name
Pagewriter Electrocardiograph
Reference
ID
RA-64204