Health Canada Medical devices

Avance and Aespire Anethiesia Machines

Issued May 30, 2022
Issue
Avance Cs2, Avance Cs2 pro anesthesia devices and field replacement batteries for Avance Cs2, Avance Cs2 pro, Avance, Amingo, Aespire View - a potential battery issue can result in premature shutdown of the anesthesia device in situations where ac mains power is lost and back up emergency power is not available. Recall start date: May 19, 2022
Hazard description
Performance
Product
Product name
Avance and Aespire Anethiesia Machines
Reference
ID
RA-64207