Back to RecallTracker
Health Canada Medical devices

Eea Autosuture Circular Stapler With Dst Series

Issued June 7, 2022
Issue
Notification to inform customers regarding potential for the staple guide to not be securely attached to the instrument. A staple guide not attached to the instrument could cause the component to disengage and if disengaged, could allow the device to transect tissue without forming staples. This could result in delay of treatment, extended hospital stay, unspecified tissue injury, unintended radiation exposure, unexpected medical intervention, foreign body in patient, failure to anastomose, and hemorrhage. Recall start date: May 16, 2022
Hazard description
Performance
Product
Product name
Eea Autosuture Circular Stapler With Dst Series
Reference
ID
RA-64233