Health Canada Medical devices

Navicare Nurse Call

Issued June 7, 2022
Issue
With certain wireless integrations, an issue has been identified when using push button call components, normal calls, staff assist, and code calls will be moved from an active state (toning and displaying on the incoming calls list) to an inactive state (no longer toning and displaying on the call history list) on the nurse call system when a call is answered at a 3rd party wireless phone with no confirmation of the audio communication between patient and caregiver, only that the caregiver answered the call. T his does not impact the initial call placement in any way and does not prevent any subsequent call from functioning properly. the requirement for the audio communication is driven by the ul 1069 standard, the regulation for hospital communication and signaling equipment. The intent is to ensure the specific patient request is communicated, as a 'routine' call may be used for multiple non urgent requests. Recall start date: May 27, 2022
Hazard description
Performance
Product
Product name
Navicare Nurse Call
Reference
ID
RA-64246