Siemens Healthcare Diagnostics Inc. has confirmed the occurrence of inconsistent discrepant (low bias) po2 results. at a medical decision level of approximately 80mmhg, the maximum negative bias is - 11.7% (the acceptable bias is +/-10% for po2 range 50-250 mmhg). Recall start date: June 8, 2022