Back to RecallTracker
Health Canada Medical devices

CT Scan and High Capacity Table

Issued March 15, 2023
Issue
During a specific servicing activity, if the wrong screws are removed by the service personnel without the appropriate table supports in place, there is the potential for the table to drop. Recall start date: June 6, 2022
Hazard description
Performance
Product
Product name
CT Scan and High Capacity Table
Reference
ID