Back to RecallTracker
Health Canada Medical devices

Vitros Immunodiagnostic Products HIV Combo Reagent Pack

Issued September 14, 2022
Issue
Scenario 1: Ortho received a customer complaint for false negative results generated from Vitros HIV combo reagent pack, lot 0670, when testing a proficiency sample identified to be reactive for the presence of HIV-1 p24 antigen and absent of antibodies to HIV (p24 antigen positive). Ortho's investigation confirmed this issue and identified five additional lots of Vitros HIV combo reagent pack that also generated false negative results when tested using the same hiv-1 p24 positive sample. For this scenario, to date, Ortho's investigation has shown that this issue is limited only to Hiv-1 p24 antigen detection and does not affect the antibody detection of antihiv-1 or anti-hiv-2 in Vitros Hiv combo reagent pack. Scenario 2: During the investigation in scenario 1, Ortho received complaints related to repeat reactive results for Vitros Hiv combo reagent pack. Ortho confirmed that known negative samples could produce reactive results. Ortho's data demonstrates a correlation to the same lots implicated in scenario 1. Recall start date: August 30, 2022
Hazard description
Performance
Product
Product name
Vitros Immunodiagnostic Products HIV Combo Reagent Pack
Reference
ID
RA-64548