Back to RecallTracker
Health Canada Medical devices

Xpert EV Assay

Issued September 20, 2022
Issue
On 27-Jul-2022 a non-conformance investigation was initiated for an issue related to a component of the test. In early August 2022, the correlation was made that this component issue may have resulted in customer complaints of invalid and false negative test results that was being investigated separately. The investigation showed that the beads used in the tests caused low sample processing control/internal control endpoints, which could lead to false negative and invalid test results. Recall start date: August 25, 2022
Hazard description
Performance
Product
Product name
Xpert EV Assay
Reference
ID
RA-64561