Aquilex Fluid Control System

Issued October 14, 2022
Issue
W.O.M. has received complaints from customers in the United States that the display of inflow volume on the Aquilex Fluid Control System can reach its limit of 30000 ml during long procedures. In these cases, the display does not indicate the actual inflow volume. Recall start date: September 21, 2022
Hazard description
Performance
Product
Product name
Aquilex Fluid Control System
Reference
ID
RA-72013