Health Canada Medical devices

Cadd Medication Cassette Reservior

Issued December 22, 2022
Issue
There is a potential that Cadd-legacy pumps may not detect that 50 ml and 100 ml Cadd medication cassette reservoirs with flow stop are attached to the pump when the cassettes are properly attached. This issue does not impact 250 ml flow-stop and non-flow stop Cadd medication cassette reservoirs. Manufacturing variations on certain cadd medication cassette reservoirs with flow stop may interfere with the pump detecting a properly attached cadd cassette. In such situations, the Cadd-legacy pump will issue a no disposable attached (NDA) double-beep audible warning if the pump cannot determine that the Cadd cassette is properly attached. The pump will initiate an NDA alarm if the nda double-beep warning is not resolved within 2 minutes. The user must clear the alarm and resolve the cause of the nda event before using the pump. Recall start date: November 29, 2022
Hazard description
Performance
Product
Product name
Cadd Medication Cassette Reservior
Reference
ID
RA-72398