Back to RecallTracker
Health Canada Medical devices

Artis Icono and Pheno Systems

Issued January 3, 2023
Issue
During transversal table movement, it could happen that the rolling bearing which is part of the linear guide rail and enables the transversal table movement may get dislocated. The rolling bearing may fall out of the linear guide rail and thus the mechanical connection between table top and table base may get lost. Recall start date: December 19, 2022
Hazard description
Device compatibility
Product
Product name
Artis Icono and Pheno Systems
Reference
ID
RA-72507