Health Canada Medical devices

Habib™ EndoHPB Catheter

Issued March 16, 2023
Issue
During investigational testing of the Habib™ Catheter with the ERBE VIO 3 generator (compatible 3rd party device), an issue with the generator setting in the Habib™ IFU (50738103) was identified. From the results of this testing, it was determined that the VIO 3 setting produces results that do not meet the dimensional requirements identified for the ablation heating zone in the Habib™ design matrix (92252607). Recall start date: March 2, 2023
Hazard description
Device compatibility - Performance
Product
Product name
Habib™ EndoHPB Catheter
Reference
ID
RA-73336