Back to RecallTracker
Health Canada Medical devices

Team3 Intrapartum and Team3 Antepartum

Issued April 13, 2023
Issue
Huntleigh Healthcare Ltd. has become aware of a software vulnerability, which manifests itself with the fetal monitor's touch screen freezing. This freeze disables the functions that are being displayed. While the occurrence of this vulnerability is extremely rare and audio remains present throughout the screen freeze, user intervention is required to restart the fetal monitor. Recall start date: April 4, 2023
Hazard description
Performance
Product
Product name
Team3 Intrapartum and Team3 Antepartum
Reference
ID
RA-73501