Back to RecallTracker
Health Canada Medical devices

Philips Respironics Trilogy EV300 and Evo ventilator

Issued April 26, 2023
Issue
Philips Respironics has detected environmental contamination (e.g. external dust and dirt) in the air path of some devices that have been returned from the field. Extended exposure to these environmental contaminants can lead to buildup of particulate on a flow sensor within the device. This may cause the device to inaccurately deliver pressure or volume/flow. Recall start date: March 28, 2023
Hazard description
Performance
Product
Product name
Philips Respironics Trilogy EV300 and Evo ventilator
Reference
ID
RA-73573