Back to RecallTracker
Health Canada Medical devices

Philips Patient Information Center

Issued May 26, 2023
Issue
During extended operation of the MX40 with the PIC iX, the DHCP (dynamic host configuration protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the pic ix, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails. Recall start date: May 12, 2023
Hazard description
Performance
Product
Product name
Philips Patient Information Center
Reference
ID
RA-73709