Health Canada Medical devices

Heartmate Touch Communication System

Issued August 18, 2023
Issue
Abbott is notifying customers of a single reported event with Heartmate Touch™ Communication System. During a clinic visit, the Heartmate Touch Communication System in one treatment room connected to the Heartmate Touch Wireless Adapter plugged into the power module in an adjacent treatment room without a user-initiated action or notification to the user. This led to the clinician adjusting settings of another patient without the clinician's awareness. Recall Start Date: August 1, 2023
Hazard description
Performance
Product
Product name
Heartmate Touch Communication System
Reference
ID
RA-74125