Back to RecallTracker
Health Canada Medical devices

Nitric Oxide Therapy Adapters And Connectors

Issued August 21, 2023
Issue
A single lot (74M2001125) of the one-way valve (part #50412 (pack of 5)) utilized with the Inomax dsir plus is being recalled. The one-way valve is used to prevent reverse flow into the Inomax dsir injector module from high frequency ventilators and spontaneous mask setups. The recall is being initiated due to the possibility of this lot of one-way valves sticking, by preventing or reducing the breathing circuit flow. We have no reports of serious injuries due to this issue. Recall Start Date: August 8, 2023
Hazard description
Performance
Product
Product name
Nitric Oxide Therapy Adapters And Connectors
Reference
ID
RA-74141