Pentacam® AXL Wave and Myopia Master®

Issued August 24, 2023
Issue
We have found an insufficient anti-reflective coating on an optical component, which can under certain circumstances lead to an additional axis length signal. If the correct axis length signal is weaker, an incorrect axis length value can be displayed. Recall start date: August 11, 2023
Hazard description
Performance
Product
Product name
Pentacam® AXL Wave and Myopia Master®
Reference
ID
RA-74195