Back to RecallTracker
Health Canada Medical devices

Elekta Infinity, Elekta Synergy and Versa Hd

Issued November 7, 2023
Issue
Elekta has identified that if the microswitch within the middle arm is incorrectly setup, there is a potential for an uncontrolled extension of iviewgt™ / xvi detector arms. when the detector arm is fully deployed there is no risk of uncontrolled extension. Recall start date: October 25, 2023
Hazard description
Performance
Product
Product name
Elekta Infinity, Elekta Synergy and Versa Hd
Reference
ID
RA-74552