Back to RecallTracker
Health Canada Medical devices

Master Vacuum Pumps

Issued November 21, 2023
Issue
When the push-start board was manufactured, a capacitor was incorrectly mounted (polarity). This can lead to overheating if the device is running for min. 60 minutes and then restarted. Delay for restart and treatment. Recall start date: November 6, 2023
Hazard description
Performance
Product
Product name
Master Vacuum Pumps
Reference
ID
RA-74621