Back to RecallTracker
Health Canada Medical devices

Defigard Touch7 Monitor And Defibrillator

Issued November 23, 2023
Issue
In rare cases, an ECG failure affecting impedance measurement via defibrillation electrodes has led to the Defigard Touch-7 falsely reports an electrode fault, resulting in inoperative defibrillation. Recall start date: November 14, 2023
Hazard description
Performance
Product
Product name
Defigard Touch7 Monitor And Defibrillator
Reference
ID
RA-74632