Back to RecallTracker
Health Canada Medical devices

Maquet Cardiopulmonary Gmbh Perfusion Sets

Issued November 28, 2023
Issue
Maquet Cardiopulmonary Gmbh received customer complaints for custom tubing packs (CTP) reporting the perforation of the Tyvek, the upper part of the sterile barrier of the packaging. Upon further investigation, the internal components set were found to be untethered/loose with the sterile tray. Recall start date: November 17, 2023
Hazard description
Sterility
Product
Product name
Maquet Cardiopulmonary Gmbh Perfusion Sets
Reference
ID
RA-74663