BBL Sensi-Disc Antimicrobial Susceptibility Test Discs

Issued January 10, 2024
Issue
BD confirmed through a recent evaluation of BD BBL™ Sensi-Disc™ product that twenty-eight (28) out of thirty (30) antimicrobial discs showed reproducibility, accuracy, and/or quality control (qc) failures when tested with Haemophilus spp. Testing with other bacterial pathogens (where indicated) is not affected. Recall Start Date: December 29, 2023
Hazard description
Device compatibility
Product
Product name
BBL Sensi-Disc Antimicrobial Susceptibility Test Discs
Reference
ID
RA-74891