Back to RecallTracker
Health Canada Medical devices

Origio Sperm Wash

Issued March 7, 2024
Issue
During testing it was found that one of the batches, had an insufficient concentration of sodium pyruvate - only 0.002 mg/ml compared to a specified concentration of 0.046-0.137 mg/ml (1.5-4.3% of the specified amount). Part of this formulation batch had also been used to create three lots of commercially released sperm wash medium. Recall start date: January 22, 2024
Hazard description
Performance
Product
Product name
Origio Sperm Wash
Reference
ID
RA-75188