Digital Radiography Mobile X-Ray System Dx-D 100

Issued March 12, 2024
Issue
A component supplier of the Agfa Dx-D 100 informed Agfa about a potential safety issue. The issue as described by the supplier is related to the steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head. The potential risk consists of an interference which could cause that the safety system which blocks the arm if the cable is cut off does not work properly. The interference occurs between the steel cable and a metal plate of the arm's carriage due to the shape of the cable's end. Recall Start Date: February 23, 2024
Hazard description
Performance
Product
Product name
Digital Radiography Mobile X-Ray System Dx-D 100
Reference
ID
RA-75221