Health Canada Medical devices

Olympus Medical Single Use Electrosurgical Knives

Issued January 6, 2026
Issue
Olympus is initiating a recall to inform customers of an increase in complaints for the triangle tip of the KD-640L and KD-645L triangle tip electrosurgical knives breaking off during use. Olympus' investigation has identified that deterioration of the cutting knife can contribute to tip breakage during use. Deterioration, including overheating and burning, may occur due to use with non-Olympus electrosurgical units and/or use of output settings that exceed the specifications.
Hazard description
Device compatibility
Product
Product name
Olympus Medical Single Use Electrosurgical Knives
Reference
ID
RA-75283