Health Canada Medical devices

Heartmate 3 LVAS Implant Kit

Issued April 5, 2024
Issue
There have been complaints of blood leaking out of the left ventricle or air entering the left ventricle or Left Ventricle Assist Device (LVAD) which are attributed to a leak path at the seal interface between the Heartmate 3 Left Ventricle Assist System (LVAS) inflow cannula and the titanium apical cuff. The blood leak or air entrainment has only been observed during the implantation procedure. Once the bleeding or air entrainment was resolved intraoperatively, the issue did not re-occur postoperatively. Abbott's investigation determined that in certain instances, routine manipulation of the pump or internal fluid pressures during implantation can result in a compressed sealing ring on one side, leading to a leak path on the opposite side. Recall Start Date: March 19, 2024
Hazard description
Performance
Product
Product name
Heartmate 3 LVAS Implant Kit
Reference
ID
RA-75348