Boston Scientific has received reports of undesired sensation and/or transient worsening of movement disorder symptoms secondary to a transient loss of stimulation during IPG charging for patients with rechargeable DBS IPGs. investigation of these reports determined that the device may reset (by design) due to detection of noise or interference during IPG charging. Recall Start Date: April 18, 2024