Health Canada Medical devices

Powerheart AED G5 Semi-Automatic and Powerheart AED G5 Semi-Automatic With CPR Assist

Issued June 17, 2024
Issue
The G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to manufacturer’s attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button. This may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest. Recall start date: June 10, 2024
Hazard description
Performance
Product
Product name
Powerheart AED G5 Semi-Automatic and Powerheart AED G5 Semi-Automatic With CPR Assist
Reference
ID
RA-75683