Back to RecallTracker
Health Canada Medical devices

Perfusor® Space® Syringe Pump

Issued June 17, 2024
Issue
Cardinal Health began distributing Monoject™ syringes branded as "Cardinal Health Monoject™ syringes" however, the new syringes have different dimensions than prior Covidien Monoject™ syringes. The dimensional changes made to the Cardinal Health Monoject™ syringes, when used with syringe pumps, may result in recognition, compatibility, and pump performance issues, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. The Perfusor® space syringe pump IFU's identify Monoject™ syringes as an option for use with the pumps. Recall start date: June 4, 2024
Hazard description
Performance
Product
Product name
Perfusor® Space® Syringe Pump
Reference
ID