Health Canada Medical devices

Centrella Max and Pro+ Mattress Replacement System (MRS) Hospital Bed Mattresses

Issued July 2, 2024
Issue
Baxter Corporation is issuing a correction for the Centrella Max and Pro+ Mattress Replacement System (MRS) Hospital Bed Mattresses due to the potential for the top cover to delaminate. Contributing causes of delamination include improper cleaning of the product (i.e., excessive use of cleaning agents containing bleach) and a lack of adequate details on the appropriate use of cleaning agents containing bleach in the product instructions for use (IFU). Recall Start Date: May 13, 2024
Hazard description
Sterility
Product
Product name
Centrella Max and Pro+ Mattress Replacement System (MRS) Hospital Bed Mattresses
Reference
ID
RA-75768