Issue
Medline Industries, LP has initiated a recall on convenience kits containing NEXUS™ Connectors A-EP-081 and A-EP-424 from Epimed International Inc. The connectors may not allow catheters to be inserted as intended if the connector is not in the fully open position. If the defect is not detected before use, there is a risk of medication leakage. Additionally, if the defect goes unnoticed, there is a risk of the catheter connector disconnecting during use. Recall start date: October 25, 2024
Hazard description
Performance