MEDRAD® Centargo Injection System

Issued January 28, 2025
Issue
Centargo patient line sensor is a component of the Centargo injection system which detects a patient line. An investigation into reported air injection concerns found that, under certain special conditions, the patient line sensor may occasionally have difficulty detecting the presence of the patient line. This intermittent detection issue, combined with user reliance on automated system functions (such as auto-prime) and possible workflow inconsistencies, may slightly increase the likelihood of small air injections (1-10 ml). However, the overall risk remains low. Recall start date: December 13, 2024
Hazard description
Performance
Product
Product name
MEDRAD® Centargo Injection System
Reference
ID
RA-76870