Health Canada Medical devices

Hamilton C2 and C3 Intensive Care Ventilators

Issued February 17, 2025
Issue
This field safety notice (FSN) provides an updated information on Hamilton medical ventilators Hamilton-C2/C3: Hamilton Medical AG became aware of a malfunction with our Hamilton-C2/C3 ventilators when used long term in neonatal patients. If an affected device Hamilton-C2 or Hamilton-C3 is used accumulatively for 91 days in neonatal patient without a restart of the device, the device will switch to "ambient state" and therefore stop ventilation. During the "ambient state" the device is alarming audibly until it is switched off by the user. Hence, it is mandatory to apply a periodic preventive restart of affected devices. Devices and software version affected: All Hamilton-C2 with sw version ≤2.2.5 and neonatal option installed All Hamilton-C3 with sw version ≤2.0.9 and neonatal option installed Recall start date: January 29, 2025
Hazard description
Performance
Product
Product name
Hamilton C2 and C3 Intensive Care Ventilators
Reference
ID