Hamilton Medical AG received several complaints where the plug (the one at the opposite side of the tubing) became detached from the In2Flow Nasal Cannula, causing leakage due to unintentional disconnection of that part. This may result in degraded high flow support and oxygen delivery, which could lead to a certain level of patient o2 desaturation. Affected product: Hamilton Medical In2Flow Nasal Cannulas with a production date dated after 31.12.2022 and till January 2025 of all sizes and all lot numbers. Recall start date: February 10, 2025