Back to RecallTracker
Health Canada Medical devices

Fujifilm Mobile Digital X-Ray System FDR Nano

Issued February 28, 2025
Issue
Fujifilm became aware that in rare cases, the arm base holding the tube section in DR-XD 1000 may get damaged. A protective hook will be attached to the equipment to prevent harm to patient even if the damage to the arm spreads. There have been no reported patient injuries associated with this issue. Recall start date: February 18, 2025
Hazard description
Performance
Product
Product name
Fujifilm Mobile Digital X-Ray System FDR Nano
Reference
ID
RA-77038