Avalon US Transducer

Issued March 12, 2025
Issue
The latest version of the wired Avalon ultrasound transducers (867246) may report inaccurate fetal heart rate (FHR) measurements when monitoring multiples (twins or triplets): in situations where the physiological signal (echo from the fetal heart) is absent or very weak (e.g. in early weeks of pregnancy), there is a tendency of the wired ultrasound transducers to interfere with each other and subsequently to produce an artificial FHR, mostly at approximately 180 bpm. The use cases affected by this issue are twin or triplet monitoring with wired ultrasound transducers including at least one transducer with software version l.01.04. The population at risk is women with multiple gestation (twins/triplets). Recall start date: March 10, 2025
Hazard description
Performance
Product
Product name
Avalon US Transducer
Reference
ID
RA-77107