CADD-Solis™ Ambulatory Infusion Pump

Issued April 28, 2025
Issue
Issue 1: Damage to the circuit board within the housing of the rechargeable battery pack may cause melting of the plastic housing, specifically on the top and bottom surfaces. Issue 2: Damage to the wireless communication module circuit board may cause melting of the plastic housing of the wireless communication module battery. Issue 3: Damaged or dislodged battery separators, or foreign material in the battery compartment, may cause an electrical short condition between the battery contacts. Recall start date: April 8, 2025
Hazard description
Performance
Product
Product name
CADD-Solis™ Ambulatory Infusion Pump
Reference
ID
RA-77345