Performing a switch device operation from the 97725 or 9772501 WENS using the A71300 SCS trialing app to transfer data (e.g. patient and lead information) to a new implanted Inceptiv or Intellis Pro Implantable Neurostimulator (INS) results in the A71400 CP app crashing and the INS becoming unusable when the transfer involves a predefined legacy lead option selected in the app that uses the forward slash character "/". Recall start date: May 28, 2025