Health Canada Medical devices

Corin Operating Tables

Issued July 24, 2025
Issue
Under certain conditions - such as the simultaneous use of the operating table with other medical devices that may cause interference, including High-Frequency (HF) surgical equipment - we have identified a potential issue that could prevent the device from performing as intended. If the issue occurs, the Corin operating table does not respond to any command. It is impossible to actuate any movement of the Corin operating table, neither with the universal remote control nor the override panel. We are not aware of instances where a patient was injured due to the described issue. Recall start date: July 18, 2025
Hazard description
Performance
Product
Product name
Corin Operating Tables
Reference
ID
RA-77760